Ibrahim, Nura Syahiera and Yusof, Firdaus and Ahmad, Norsham (2020) The consistency of retinal image size measurement using smartphone application. Malaysian Journal of Medicine and Health Sciences (MJMHS), 16 (3). pp. 15-20. E-ISSN 2636-9346

[img]
Preview
PDF - Published Version
Download (489kB) | Preview
[img]
Preview
PDF (SCOPUS) - Supplemental Material
Download (201kB) | Preview

Abstract

Introduction: Unequal retinal image size (RIS) or aniseikonia is usually related with anisometropia. Higher differences of RIS may manifest symptoms such as dizziness, headache or disorientation. In worst case might cause suppression that leads to amblyopia. Current study aims to evaluate the consistency of aniseikonia measurement in Smart Optometry smartphone application among myopic, hyperopic, and astigmatic simulated anisometropia and real anisometropia groups. Methods: Fifteen real anisometropes (refractive error; -0.50 until -6.00 diopters; D) and fifteen emmetropes (refractive error: -0.25 until +0.50D) were recruited. Real anisometropes wore their habitual spectacle correction while each emmetropes were fitted using soft contact lenses of +4.00DS, -4.00DS and -4.00DC with base curve 8.6 and total diameter 14.2mm in random order to mimic myopic-, hyperopic- and astigmatic-anisometropia before testing. Participants with any ocular disease and binocular vision problem were excluded. The consistency of aniseikonia measurement was determined in two visits, separated by at least 24-hour interval. Three repetitive measurements were taken in each visit. Results: Independent t-test and paired t-test showed that real and simulated anisometropia gave insignificant aniseikonia percentage, p>0.05. ICC findings revealed moderate-to-good agreement for all simulated and real groups. Bland Altman analysis between two visits exhibited good agreement among all simulated group; myopic (mean difference 0.2047; 95%CI:-1.1386-1.549), hyperopic (mean difference 0.2200; 95%CI:-0.9286-1.3686) and astigmatic (mean difference 0.2533; 95%CI:-0.7114-1.2180). Real anisometropes demonstrated good agreement with bias value of 0.2247(95%CI:-0.9162-1.3656) using Bland Altman plot. Conclusion: Smart Optometry application provides consistent measurement of aniseikonia regardless any types of anisometropia

Item Type: Article (Journal)
Additional Information: 4598/83148
Uncontrolled Keywords: Aniseikonia, Anisometropia, Simulated anisometropia
Subjects: R Medicine > RE Ophthalmology
Kulliyyahs/Centres/Divisions/Institutes (Can select more than one option. Press CONTROL button): Kulliyyah of Allied Health Sciences > Department of Optometry and Visual Science
Kulliyyah of Allied Health Sciences
Depositing User: Dr Norsham Ahmad
Date Deposited: 09 Oct 2020 09:04
Last Modified: 12 Nov 2020 11:51
URI: http://irep.iium.edu.my/id/eprint/83148

Actions (login required)

View Item View Item

Downloads

Downloads per month over past year