Image quality and entrance surface dose evaluation for lateral cervical spine: a study using grid and non-grid techniques

Moey, Soo Foon and Mohd Ramlee, Nurfatin Nabilah (2018) Image quality and entrance surface dose evaluation for lateral cervical spine: a study using grid and non-grid techniques. Iranian Journal of Medical Physics, 16 (2 (March-April)). pp. 166-170. ISSN 2345-3672

[img] PDF - Published Version
Restricted to Registered users only

Download (839kB) | Request a copy
[img]
Preview
PDF (SCOPUS) - Supplemental Material
Download (188kB) | Preview

Abstract

Introduction: The purpose of this study is to ascertain the effect of a moving grid, stationary grid and non-grid in the lateral cervical spine radiography on image quality and entrance surface dose (ESD). Image quality and radiation doses have been studied by various authors, but there is limited amount of information available on image quality and patient dose from cervical spine examination with different grid techniques. Materials and Methods: The radiographs of the lateral cervical spine were acquired by positioning the RANDO phantom abutting the erect bucky whilst using a moving grid, stationary grid and non-grid. A 24 cm x 30 cm Fuji standard cassette type imaging plate was used in this study. A Leeds TOR test tool was utilized for relative comparison of image quality for the fore mentioned techniques. The ESD of each examination was determined by using the Optically Stimulated Luminescence Dosimeter (OSLD). Results: Increased kilovoltage (kVp) resulted in the reduction of ESD whether moving grid, stationary grid or non-grid was utilized. Significant differences in high contrast sensitivity, low contrast sensitivity and spatial resolution were reflected when using the grid compared to that of non-grid, x2(2, N=42) = 8.000; 5.000, p<0.05. The results also indicated significant differences in ESD when using a moving grid, stationary grid and non-grid, x2(2, N=42) = 7.200, p<0.05. Conclusion: This study reflected that there were significant differences in image quality and ESD when a moving grid, stationary grid and non-grid were used in lateral cervical spine radiography. Further, non-grid cervical spine lateral radiography using the highest appropriate kVp is recommended as the existed air gap acts as a grid, resulting in acceptable image quality with reduction in ESD.

Item Type: Article (Journal)
Additional Information: 7455/66537
Uncontrolled Keywords: Digital radiography, image quality, radiation dosage, cervical vertebrae
Subjects: Q Science > QC Physics
R Medicine > R Medicine (General)
Kulliyyahs/Centres/Divisions/Institutes (Can select more than one option. Press CONTROL button): Kulliyyah of Allied Health Sciences
Kulliyyah of Allied Health Sciences > Department of Diagnostic Imaging and Radiotherapy
Depositing User: Dr Soo Foon Moey
Date Deposited: 11 Oct 2018 14:51
Last Modified: 15 Mar 2019 09:12
URI: http://irep.iium.edu.my/id/eprint/66537

Actions (login required)

View Item View Item

Downloads

Downloads per month over past year

Loading...