Moey, Soo Foon and Zubir, Ahmad Shazli and Sayed, Inayatullah Shah (2018) The practice of chest radiography using different digital imaging systems: Dose and image quality. Iranian Journal of Medical Physics, 15 (2). pp. 101-107. E-ISSN 2345-3672
PDF
- Published Version
Restricted to Registered users only Download (373kB) | Request a copy |
|
PDF (SCOPUS)
- Supplemental Material
Restricted to Registered users only Download (526kB) | Request a copy |
Abstract
Introduction: The study was undertaken to evaluate the practice of chest radiography using different digital imaging systems and its influence on dose and image quality. Materials and Methods: The study was carried out in two hospitals from March 2016 to June 2016. Sixty ambulatory patients aged 21 to 60 years who were able to cooperate without difficulty and weighed between 60 to 80 kg were selected randomly. The active matrix flat panel imagers technology was employed in the direct radiography (DR) system for Hospital A, whilst Hospital B used the single read out computed radiography (CR) system. The dose area product (DAP) meter was utilized in measuring the entrance surface air kerma. The chest radiographs were evaluated by two radiologists. Results: The mean entrance surface doses (ESDs) for posteroanterior chest in Hospital A (0.098 mGy) was lower than that obtained in Hospital B (0.161 mGy). However, the ESDs at both centres were lower than the recommended value by the International Atomic Energy Agency (IAEA; 0.3 mGy). The quality of the images for chest radiography in both hospitals was adequate to make a diagnosis with ESDs and effective doses lower than those recommended by IAEA and United Nations Scientific Committee on the Effects of Atomic Radiation (UNSCEAR). Conclusion: The study serves to highlight the practice of chest radiography with two different systems and its influence on image quality and dose. It can be concluded that there were significant differences in image quality and radiation dose for chest radiography practice using CR and DR.
Item Type: | Article (Journal) |
---|---|
Additional Information: | 7455/63147 |
Uncontrolled Keywords: | Digital Radiography,Image Quality,Radiation Dosage,Thoracic Radiography |
Subjects: | Q Science > QC Physics R Medicine > R Medicine (General) |
Kulliyyahs/Centres/Divisions/Institutes (Can select more than one option. Press CONTROL button): | Kulliyyah of Allied Health Sciences > Department of Diagnostic Imaging and Radiotherapy |
Depositing User: | Dr Soo Foon Moey |
Date Deposited: | 11 Apr 2018 15:04 |
Last Modified: | 11 Apr 2018 15:04 |
URI: | http://irep.iium.edu.my/id/eprint/63147 |
Actions (login required)
View Item |