Abu-Samah, Asma and Knopp, Jennifer Launa and Abdul Razak, Normy Norfiza and Abdul Razak, Athirah and Jamaludin, Ummu Kulthum and Mohamad Suhaimi, Fatanah and Md Ralib, Azrina and Mat Nor, Mohd Basri and Chase, James Geoffrey and Grant Pretty, Christopher (2019) Model-based glycemic control in a Malaysian intensive care unit: performance and safety study. Medical Devices: Evidence and Research, 2019. pp. 215-226. ISSN 1179-1470

[img]
Preview
PDF - Published Version
Download (2MB) | Preview
[img]
Preview
PDF (Scopus) - Supplemental Material
Download (792kB) | Preview
[img]
Preview
PDF (wos) - Supplemental Material
Download (422kB) | Preview

Abstract

Background: Stress-induced hyperglycemia is common in critically ill patients. A few forms of model-based glycemic control have been introduced to reduce this phenomena and among them is the automated STAR protocol which has been used in the Christchurch and Gyulá hospitals’ intensive care units (ICUs) since 2010. Methods: This article presents the pilot trial assessment of STAR protocol which has been implemented in the International Islamic University Malaysia Medical Centre (IIUMMC) Hospital ICU since December 2017. One hundred and forty-two patients who received STAR treatment for more than 20 hours were used in the assessment. The initial results are presented to discuss the ability to adopt and adapt the model-based control framework in a Malaysian environment by analyzing its performance and safety. Results: Overall, 60.7% of blood glucose measurements were in the target band. Only 0.78% and 0.02% of cohort measurements were below 4.0 mmol/L and 2.2 mmol/L (the limitsfor mild and severe hypoglycemia, respectively). Treatment preference-wise, the clinical staff were favorable of longer intervention options when available. However, 1 hourly treatments were still used in 73.7% of cases. Conclusion: The protocol succeeded in achieving patient-specific glycemic control while maintaining safety and was trusted by nurses to reduce workload. Its lower performance results, however, give the indication for modification in some of the control settings to better fit the Malaysian environment. © 2019 Abu-Samah et al.

Item Type: Article (Journal)
Additional Information: 5608/76338
Uncontrolled Keywords: Glycemic control, Intensive care unit, Malaysian hospital, Model -based control, Pilot trial
Subjects: R Medicine > R Medicine (General)
R Medicine > RZ Other systems of medicine
Kulliyyahs/Centres/Divisions/Institutes (Can select more than one option. Press CONTROL button): Kulliyyah of Medicine
Kulliyyah of Medicine > Department of Anaesthesiology & Intensive Care
Depositing User: Dr. Mohd Basri Mat Nor
Date Deposited: 18 Dec 2019 15:00
Last Modified: 17 Jul 2020 09:20
URI: http://irep.iium.edu.my/id/eprint/76338

Actions (login required)

View Item View Item

Downloads

Downloads per month over past year